• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. HIGH FLOW INSUFFLATION UNIT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. HIGH FLOW INSUFFLATION UNIT Back to Search Results
Model Number UHI-4
Device Problem Device Contamination with Chemical or Other Material (2944)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/10/2021
Event Type  malfunction  
Manufacturer Narrative
Investigation activities have been opened to manage the actions related to this issue and any required mdr reporting.The device is returned and an evaluation completed for it.The manufacture date is not available.The user¿s complaint was confirmed.Upon inspection and testing, unknown liquid was observed inside pneumatic system tubes, manifold unit, electro-pneumatic proportional valve, first regulator unit and k-connector unit.All these parts would need to be replaced.Evaluation is ongoing.Supplemental report(s) will be submitted when any relevant new information is available.
 
Event Description
As reported for this event by the customer, during a preventive maintenance unknown liquid was detected in the electro-pneumatic hoses.There is no patient involvement and no reported harm to any patient.
 
Manufacturer Narrative
There is more information on the device evaluation.This supplemental report is being submitted to provide this information.Device history record review indicates that the device was manufactured and tested in accordance with all applicable procedures and met all final product release criteria.The root cause for the issue cannot be conclusively determined.However, it is likely that the foreign body (fluid marks) in the entire tract resulted from reflux from the gas cylinder or the tube side.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIGH FLOW INSUFFLATION UNIT
Type of Device
HIGH FLOW INSUFFLATION UNIT
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
MDR Report Key12534217
MDR Text Key273557073
Report Number8010047-2021-12343
Device Sequence Number1
Product Code HIF
UDI-Device Identifier04953170324147
UDI-Public04953170324147
Combination Product (y/n)N
PMA/PMN Number
K122180
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,user facility
Type of Report Initial,Followup
Report Date 10/27/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/27/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberUHI-4
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/30/2021
Was the Report Sent to FDA? No
Date Manufacturer Received10/25/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-