After further reviewed of additional information received the following sections have been updated accordingly :g3, g4, g6, h1, h2, h3 and h6.As reported a si brite tip 6f 11cm str was used for an endovascular therapy case.The device was noted to be ¿turned up¿ and a crack was confirmed at the distal end of the device.The device could not be used for the procedure.There was no reported patient injury.The device was not returned for analysis.A product history record (phr) review of lot 18024811 revealed no anomalies or non-conformances during the manufacturing and inspection processes that can be associated with the reported event.The reported ¿brite tip/distal tip-cracked - during prep¿ could not be confirmed as the device was not returned for analysis.The exact cause of the event could not be determined.Procedural factors, such as operator handling during prep, could have contributed to the event.According to the instructions for use (ifu) which is not intended as a mitigation of risk, ¿do not use if package is open or damaged.¿ users are clinically trained to inspect devices prior to use and during prep and immediate discontinue and change the device if any damage is noted.Neither the phr, nor the information available, suggests a design or manufacturing related cause for the reported event; therefore, no corrective/preventive action will be taken at this time.
|