DEPUY IRELAND - 9616671 ATTUNE DISTAL FEM CUT BLOCK; ATTUNE INSTRUMENTS : RESECTION BLOCKS/DEVICES Back to Search Results
Model Number 2544-00-522
Device Problem Appropriate Term/Code Not Available (3191)
Patient Problem Insufficient Information (4580)
Event Date 08/03/2021
Event Type  malfunction  
Manufacturer Narrative
Product complaint #: (b)(4).Investigation summary: examination of the returned cutting block identified the presence of a burr around the edges of the cutting slot.The noted damage is consistent with device wear out due to normal intended use and the investigation did not establish a need for corrective action.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that the doctor questioned the tolerances of how much it was going to resect.Knob broke off of fem introducer while impacting.There was 5 minutes surgical delay.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE DISTAL FEM CUT BLOCK
Type of Device
ATTUNE INSTRUMENTS : RESECTION BLOCKS/DEVICES
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46581 0988
Manufacturer Contact
kara ditty-bovard
700 orthopaedic dr.
warsaw, IN 46581-0988
6107428552
MDR Report Key12541847
MDR Text Key273627809
Report Number1818910-2021-21302
Device Sequence Number1
Product Code HTZ
UDI-Device Identifier10603295423737
UDI-Public10603295423737
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 08/03/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/28/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number2544-00-522
Device Catalogue Number254400522
Device Lot NumberABB35809
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/16/2021
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/27/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/19/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age68 YR
-
-