DATASCOPE CORP. - MAHWAH CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 110V SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number 0998-00-3023-53
Device Problem Infusion or Flow Problem (2964)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/05/2021
Event Type  malfunction  
Manufacturer Narrative
Testing of actual/suspected device: a getinge field service engineer (fse) was dispatched to evaluate this unit. Fse installed 2. 5 kit on compressor to ensure proper operation. United failed k6a,k6,k7,k8 test indicating signs of failed k8 valve, so fse replaced drive manifold. Fse completed all safety, functionality, and calibration checks and all tests passed to factory specifications. The iabp unit was cleared for clinical use and released to the customer. A supplemental report will be submitted if additional information is provided.
 
Event Description
It was reported that during use the cs300 intra-aortic balloon pump (iabp) had low vacuum alarms. Customer switched consoles, took only seconds of therapy delay. There was no harm or injury to the patient and no adverse event was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameCS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 110V
Type of DeviceSYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
dorota wolpiuk
1300 macarthur blvd
mahwah, NJ 
MDR Report Key12541957
MDR Text Key273632794
Report Number2249723-2021-02186
Device Sequence Number1
Product Code DSP
Combination Product (y/n)N
PMA/PMN Number
K063525
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation
Type of Report Initial
Report Date 09/28/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/28/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number0998-00-3023-53
Device Catalogue Number0998-00-3023-53
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received09/05/2021
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device?
Type of Device Usage Reuse

Patient Treatment Data
Date Received: 09/28/2021 Patient Sequence Number: 1
Treatment
MAQUET BALLOON, TYPE UNKNOWN.
-
-