According to the reporter, during procedure, the balloon did not pass through a previously deployed stent; it was utilized with a 035 guide wire (bard), and resistance encountered when advancing it during first attempt.
The balloon was inflated with syringe, no inflation fluid was used and it was deflated normally.
After the anastomosis was cleared in the central artery, during removal/withdrawal of balloon/wire from the patient, excessive/additional force was used as there was some scare tissue in surrounding entrance site and the catheter broke/snapped at the side hole (exit port).
This was a sheath less procedure.
The doctor had to cut down and use snare to remove the balloon.
No additional anesthesia required to perform the cutdown.
Additional surgical time was needed.
A diagnostic testing (fluoroscopy) was done as a part of procedure to verify if there were no parts retained on the patient¿s body.
All pieces were accounted for and removed on the same day of the event.
Nothing unusual observed on the device prior to use.
The device was flushed with no issues identified.
The catheter was not repaired, there was no leak, tego was not utilized, and there was no luer adapter issue.
The insertion site was not treated with prior to product placement.
Saline flush was the cleaning agent used on the device.
There was no blood loss.
Blood transfusion was not required.
No additional medical treatments or medications.
No hospital admission or prolonged hospitalization.
The patient did not experience any symptoms and was in good condition.
|