Model Number X SERIES |
Device Problem
Self-Activation or Keying (1557)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/08/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
|
|
Event Description
|
Complainant alleged that while attempting to monitor a patient (age & gender unknown), the device self-charged without any user interaction.Complainant indicated that there was no adverse effect to the patient due to the reported malfunction.
|
|
Manufacturer Narrative
|
Zoll medical corporation evaluated the device and the device performed to specification.The device was recertified and returned to the customer.Review of the device activity logs showed no evidence of the customer's report.The log showed that the device operated in manual mode throughout the case.Therefore, any analysis events would require to be manually initiated.The log showed no clinical analysis or charge events.No critical errors were found in the logs.The customer's report was unsubstantiated.Analysis of reports of this type has not identified an increase in trend.
|
|