• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW MEDICAL LTD. RENASYS TOUCH NON CONNECT 4TH ED DEVICE; NEGATIVE PRESSURE WOUND THERAPY POWERED SUCTION PUMP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW MEDICAL LTD. RENASYS TOUCH NON CONNECT 4TH ED DEVICE; NEGATIVE PRESSURE WOUND THERAPY POWERED SUCTION PUMP Back to Search Results
Catalog Number 66802134
Device Problems Fire (1245); Overheating of Device (1437)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/09/2021
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that the renasys touch was found on fire.No case involved; therefore, there was no patient involvement.
 
Manufacturer Narrative
Updated d9, h6.H10: the device, intended for use in treatment, has been returned and evaluated with photographic evidence.This reported event has been assessed in detail.On cross examination with another device where internal combustion was evident; the battery tops had vented showing evidence of defective batteries, the circuit board was mostly intact with residue of battery vent/discharge and evidence of smoke which exited from within a contained environment, and reducing external damage with no single burn points.In regard to this reported event, the battery tops were intact, the circuit board had disintegrated showing evidence that the fire was not contained within the device.There is a single burn point to the outer casing, with evidence of extensive external burning, possibly due to the device burning for a longer period of time.No evidence of battery venting was found.Conclusion: a visual inspection reported extensive external and internal burn damage to multiple components, establishing a relationship between the reported event.All evidence suggests that the probable root cause is due to misuse of the device.The ifu provides safe use and handling and storage.Keep the device away from direct heat sources during charging.The associated risk files have been reviewed and require no updates.A review of the manufacturing records found that there was no evidence that the product didn¿t meet specifications at the time of manufacture.A complaint history review found a further instance of this event type.This investigation is now complete with no further action deemed necessary.Smith + nephew will continue to monitor for any adverse trends relating to this product range.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RENASYS TOUCH NON CONNECT 4TH ED DEVICE
Type of Device
NEGATIVE PRESSURE WOUND THERAPY POWERED SUCTION PUMP
Manufacturer (Section D)
SMITH & NEPHEW MEDICAL LTD.
101 hessle road
hull east riding of yorkshire HU3 2 BN
UK  HU3 2BN
Manufacturer (Section G)
SMITH & NEPHEW MEDICAL LTD.
101 hessle road
hull east riding of yorkshire HU3 2 BN
UK   HU3 2BN
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key12550976
MDR Text Key273971918
Report Number8043484-2021-01812
Device Sequence Number1
Product Code OMP
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K181204
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 11/24/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/29/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number66802134
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/29/2021
Is the Reporter a Health Professional? No
Date Manufacturer Received11/04/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-