• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DATASCOPE CORP. - FAIRFIELD LINEAR 7.5 FR. 40CC IAB; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE CORP. - FAIRFIELD LINEAR 7.5 FR. 40CC IAB; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number 0684-00-0480-01
Device Problem Material Deformation (2976)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/09/2021
Event Type  malfunction  
Manufacturer Narrative
The initial reporter named in is a getinge employee whose contact details are: (b)(6) ; which differs from that of the event site.The product has been returned to the manufacturer, but is pending investigation.Once the investigation is completed a supplemental report with our findings will be submitted.Complaint record id # (b)(4).
 
Event Description
During the investigation of the returned intra-aortic balloon (iab) involved under mfg report number 2248146-2021-00356, a kink was found that was unrelated to the reported silicone migration (condensation) failure mode.There was no patient involvement.
 
Manufacturer Narrative
Only the packaging components were returned and unused.The packaging was not returned.A catheter tubing kink was found at 27.94 cm from the iab tip.The evaluation confirms the presence of a catheter tubing kink as an as analyzed failure.However, we are unable to conclusively determine when this kink may have occurred.Therefore, the root cause for the kink is impossible to define.A device and lot history record review was completed for the reported product.No nonconformances were found that are considered to be related to the event.Analysis of production (3331/213) - the device history records review concluded that there were no ncmrs, rework, or deviations documented for the reported lot/serial number.Based on the review results, it was determined that there is no relation between the manufacturing process and the reported failure.Historical data analysis (4109/213) - the review of the historical data was performed.Trend analysis (4110/213) - the overall complaint trend data for the period aug-2019 through jul-2021 was reviewed.Communication/interviews: (4111/213) communication/interviews were performed to obtain all possible information.Reference complaint #(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LINEAR 7.5 FR. 40CC IAB
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - FAIRFIELD
15 law drive
fairfield NJ
MDR Report Key12567927
MDR Text Key274569166
Report Number2248146-2021-00643
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567106564
UDI-Public10607567106564
Combination Product (y/n)N
PMA/PMN Number
K041281
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 10/13/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/04/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/10/2022
Device Model Number0684-00-0480-01
Device Catalogue Number0684-00-0475
Device Lot Number3000106389
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/13/2021
Date Manufacturer Received10/13/2021
Patient Sequence Number1
-
-