• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID, TYPE B PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID, TYPE B PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number 0998-00-0800-53
Device Problem Unexpected Shutdown (4019)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/10/2021
Event Type  malfunction  
Manufacturer Narrative
Testing of actual/suspected device : a getinge field service engineer (fse) evaluated the iabp, but was unable to reproduce the reported issue.The fse turned the iabp unit on to make sure it auto filled and pumped, checked logs and was not able to find any errors.He then performed all functional and safety checks to meet factory specifications.Unit passed all functional and safety test per factory specifications.The iabp was then released to the customer and cleared for clinical service.A supplemental report will be submitted upon completion of our investigation.The full name of the initial reporter that is abbreviated is (b)(6).The full event site name in is (b)(6).
 
Event Description
It was reported that during patient transport the cardiosave intra-aortic balloon pump (iabp) failed.It was later clarified that the iabp unit had shut off.No patient harm, serious injury or adverse event was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOSAVE HYBRID, TYPE B PLUG
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key12572850
MDR Text Key274760831
Report Number2249723-2021-02247
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567108391
UDI-Public10607567108391
Combination Product (y/n)N
PMA/PMN Number
K112372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 09/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/04/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0998-00-0800-53
Device Catalogue Number0998-00-0800-53
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received09/20/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/27/2018
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-