On 09/17/2021, it was reported by an arthrex employee via sems that a set of ar-6410 main pump tubing did not work, causing the console to display "door open".
This was discovered during use in a procedure performed on (b)(6) 2021.
The case was completed by using gravity-assisted-pumping techniques with no patient effect reported.
Update: case is confirmed to be a significant deviation from procedure consistent with aborted procedure - serious injury.
Information on pump requested from re and will be entered as second product line when received.
Update: contact with rep continued.
Rep claimed the issue resulted from a setting mistake when attempting to use the pump and tubing, and the procedure was completed using another device.
The part and serial numbers of the pump used could not be identified.
Due to lack of complete information, case will be left as reportable, subject to change if more information is obtained.
|