BD MEDICAL (BD WEST) MEDICAL SURGICAL BD LUER-LOK TIP SYRINGE PISTON SYRINGE Back to Search Results
Model Number 309653
Device Problem Leak/Splash (1354)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/07/2021
Event Type  malfunction  
Manufacturer Narrative
Date of event: unknown. The date received by manufacturer has been used for this field. Medical device expiration date: unknown. Device manufacture date: unknown. Investigation summary: as no physical sample, picture sample, or lot number was provided for evaluation by our quality engineer team, a complete investigation could not be performed. There are current quality controls in place to detect this type of product malfunction during the production process. Based on the limited investigation results, a cause for the reported incident could not be determined. Complaints received for this device and reported condition will continue to be tracked and trended. Our quality team regularly reviews the collected data for identification of emerging trends.
 
Event Description
It was reported that the bd luer-lok¿ tip syringe leaked a water-based drug when used with the harvard apparatus 33 dds syringe pump. The following information was provided by the initial reporter: "states they are using the syringe with a harvard apparatus 33 dds syringe pump with 1/16 micro tubing and are experiencing leaking. States the educational drug is water-based and they are filling the syringe with 20ml of fluid volume. They do not have a product code or reference number for the tubing they are using with the syringe. ".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameBD LUER-LOK TIP SYRINGE
Type of DevicePISTON SYRINGE
Manufacturer (Section D)
BD MEDICAL (BD WEST) MEDICAL SURGICAL
1852 10th avenue
columbus NE 68601
Manufacturer (Section G)
BD MEDICAL (BD WEST) MEDICAL SURGICAL
1852 10th avenue
columbus NE 68601
Manufacturer Contact
katie swenson
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key12579582
MDR Text Key275040065
Report Number1911916-2021-01057
Device Sequence Number1
Product Code FMF
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K110771
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other,user facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 09/17/2021
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received10/05/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number309653
Device Catalogue Number309653
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received09/07/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-