Catalog Number 828806
Device Problem Infusion or Flow Problem (2964)
Patient Problem Failure of Implant (1924)
Event Type  Injury  
Manufacturer Narrative
The certas valve was not returned for evaluation (remains implanted) and no lot number information has been provided; therefore, an evaluation of the device could not be performed, and manufacturing records could not be reviewed. The cause(s) of the difficulty reported by the customer could not be determined. If additional relevant information becomes available in the future, this complaint will be reopened, and the respective evaluation performed. Trends will be monitored for this and similar issues. At present, we consider this complaint to be closed.
 
Event Description
A physician reported a certas valve was implanted in a patient due to inph (idiopathic normal pressure hydrocephalus) via l-p shunt on unknown date with unknown setting. The valve was used with the silascon lumbar catheter (manufactured by kaneka, product code: 702-jj). Underdrainage occurred, and rupture was confirmed through contrast and imaging. The rupture was at the connection between the stepped connector and the lumbar catheter. The lumbar catheter was pulled out slightly, the torn part was cut, and the ruptured part was reconnected to the stepped connector to complete the procedure. It is unknown if the patient experienced any signs and symptoms.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameCERTAS INLIN VLV SPHN/UNIT CAT
Type of DeviceCERTAS PLUS W/ SG
Manufacturer (Section D)
INTEGRA LIFESCIENCES SWITZERLAND SAR
rue girardet 29
rue girardet 29
le locle
SZ
Manufacturer (Section G)
INTEGRA LIFESCIENCES SWITZERLAND SAR
rue girardet 29
ch-2400
le locle
SZ
Manufacturer Contact
vivian nelson
1100 campus drive
princeton, NJ 
6099362319
MDR Report Key12582771
MDR Text Key274923462
Report Number3013886523-2021-00424
Device Sequence Number1
Product Code JXG
Combination Product (y/n)N
PMA/PMN Number
K143111
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Reporter Occupation
Type of Report Initial
Report Date 10/06/2021
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received10/06/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Device Operator
Device Catalogue Number828806
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received09/16/2021
Was Device Evaluated by Manufacturer? No Answer Provided
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unkown

-
-