• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DATASCOPE CORP. - FAIRFIELD SENSATION PLUS 8FR. 50CC IAB SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE CORP. - FAIRFIELD SENSATION PLUS 8FR. 50CC IAB SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number 0684-00-0576-01
Device Problem Inappropriate Waveform (2536)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/17/2021
Event Type  malfunction  
Manufacturer Narrative
Complete initial reporter name: (b)(6). The device has not been returned to the manufacturer so we are unable to complete an evaluation. If provided we will send a supplemental report with our additional findings. Complaint record id # (b)(4).
 
Event Description
It was reported that during intra-aortic balloon (iab) therapy, it was noted that the arterial waveform was full of artifact. The physician in the post-op surgical unit insisted the tip was the right position. The customer was assisted in substituting the radial arterial line as an alternate arterial source. There was no patient harm or adverse event reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameSENSATION PLUS 8FR. 50CC IAB
Type of DeviceSYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - FAIRFIELD
15 law drive
fairfield NJ
Manufacturer (Section G)
DATASCOPE CORP. - FAIRFIELD
15 law drive
fairfield NJ
Manufacturer Contact
brian schaeffer
15 law drive
fairfield, NJ 
MDR Report Key12586695
MDR Text Key275195819
Report Number2248146-2021-00657
Device Sequence Number1
Product Code DSP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K112327
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation
Type of Report Initial,Followup
Report Date 11/11/2021
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received10/06/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number0684-00-0576-01
Device Catalogue Number0684-00-0575
Device Lot Number3000140637
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/11/2021
Was Device Evaluated by Manufacturer? No
Date Device Manufactured01/08/2021
Is the Device Single Use? Yes
Type of Device Usage Initial

-
-