• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) CONFIRM; IMPLANTABLE CARDIAC MONITOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) CONFIRM; IMPLANTABLE CARDIAC MONITOR Back to Search Results
Model Number DM3500
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Erosion (1750)
Event Date 09/28/2021
Event Type  Injury  
Event Description
It was reported that the patient's implantable cardiac monitor (icm) was working it's way out of the incision following implant.Erosion was confirmed.It was believed the insertion site had not been closed properly.The icm was explanted and replaced.The patient was stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONFIRM
Type of Device
IMPLANTABLE CARDIAC MONITOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
sonali arangil
15900 valley view court
sylmar, CA 91342
MDR Report Key12594446
MDR Text Key275378222
Report Number2017865-2021-32829
Device Sequence Number1
Product Code MXC
UDI-Device Identifier05415067027320
UDI-Public05415067027320
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K163407
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 10/07/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/07/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date09/06/2021
Device Model NumberDM3500
Device Catalogue NumberDM3500
Device Lot NumberP000100790
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received09/28/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/12/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age72 YR
-
-