Model Number 0998-00-0800-53
Device Problem Failure to Charge (1085)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/15/2021
Event Type  malfunction  
Manufacturer Narrative
Testing of actual/suspected device : a getinge field service engineer (fse) evaluated the iabp unit for the reported "battery not charging after being replaced", he troubleshot the system and determined that the power management board was faulty. To fix the issue, he replaced the power management board and confirmed that the batteries were charging. He left the unit connected to the a/c power overnight to allow the batteries to fully charge prior to completing the functionality and operational checks. The fse returned on site and confirmed batteries were fully charged and performed all functional and safety checks to meet factory specifications. Unit passed all functional and safety test per factory specifications. The iabp was then released to the customer and cleared for clinical service. A supplemental report will be submitted upon completion of our investigation. The full event site name in is (b)(6). The full name of the initial reporter that is abbreviated in is (b)(6).
 
Event Description
It was reported that prior to use the battery on the cardiosave intra-aortic balloon pump (iabp) would not charge after it was replaced. There was no patient involvement, and no adverse event reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameCARDIOSAVE HYBRID, TYPE B PLUG
Type of DeviceSYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
dorota wolpiuk
1300 macarthur blvd
mahwah, NJ 
MDR Report Key12597755
MDR Text Key275412739
Report Number2249723-2021-02288
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567108391
UDI-Public10607567108391
Combination Product (y/n)N
PMA/PMN Number
K112372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 10/07/2021
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received10/08/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number0998-00-0800-53
Device Catalogue Number0998-00-0800-53
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received09/15/2021
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device?
Type of Device Usage Reuse

-
-