Model Number AED3
Device Problem Failure of Device to Self-Test (2937)
Patient Problem Insufficient Information (4580)
Event Type  malfunction  
Manufacturer Narrative
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
 
Event Description
Complainant alleged that during functional testing, the device prompted a "unit failed" message.Complainant indicated that there was no patient involvement in the reported malfunction.
 
Manufacturer Narrative
The device was returned to zoll medical corporation.The customer's report was not replicated or confirmed.The device was put through extensive testing including bench handling, power cycling, functional stress testing, and on/off current testing using test batteries and pads without duplicating the report.The device was recertified and returned to the customer.The battery used was not returned as part of this investigation.Analysis of reports of this type has not identified an increase in trend.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AED3
Type of Device
DEFIBRILLATOR
Manufacturer (Section D)
ZOLL MEDICAL CORPORATION
269 mill road
chelmsford MA 01824
Manufacturer Contact
269 mill road
chelmsford, MA 01824
9784219552
MDR Report Key12609785
MDR Text Key277378654
Report Number1220908-2021-03333
Device Sequence Number1
Product Code MKJ
UDI-Device Identifier00847946020071
UDI-Public00847946020071
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P160022
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 09/20/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/11/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberAED3
Device Catalogue NumberAED3
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/06/2021
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received09/20/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/01/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-