• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BD MEDICAL - DIABETES CARE SYRINGE 0.5ML 31GA 8MM UFII 10BAG 500CS; PISTON SYRINGE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BD MEDICAL - DIABETES CARE SYRINGE 0.5ML 31GA 8MM UFII 10BAG 500CS; PISTON SYRINGE Back to Search Results
Catalog Number 328468
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/14/2021
Event Type  malfunction  
Manufacturer Narrative
Date of event: unknown.The date received by manufacturer has been used for this field.Investigation summary: samples were received and an investigation was performed.This is the 1st complaint for the reported lot number.A review of the manufacturing records was performed and no non-conformances were raised in association with this type of event for this lot.Bd was able to duplicate or confirm the indicated issue and based on trend analysis no further action is required at this time.Complaints received for this device and reported condition will continue to be tracked and trended.Based on the above, no additional investigation and no corrective/preventative action (capa) or situational analysis (sa) is required at this time.
 
Event Description
It was reported that syringe 0.5ml 31ga 8mm ufii 10bag 500cs was cracked.The following information was provided by the initial reporter: it was reported by the consumer, the barrel of the syringe has a crack near the end where the needle is.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYRINGE 0.5ML 31GA 8MM UFII 10BAG 500CS
Type of Device
PISTON SYRINGE
Manufacturer (Section D)
BD MEDICAL - DIABETES CARE
1329 west highway 6
holdrege NE 68949
Manufacturer (Section G)
BD MEDICAL - DIABETES CARE
1329 west highway 6
holdrege NE 68949
Manufacturer Contact
katie swenson
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key12615942
MDR Text Key275953405
Report Number1920898-2021-01068
Device Sequence Number1
Product Code FMF
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K024112
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,other
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 10/01/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/12/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue Number328468
Device Lot Number1130863
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/04/2021
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/14/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/10/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-