BOSTON SCIENTIFIC CORPORATION INVIVE; IMPLANTABLE DEVICE Back to Search Results
Model Number V173
Device Problem Incorrect, Inadequate or Imprecise Result or Readings (1535)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/26/2021
Event Type  Injury  
Manufacturer Narrative
Information indicates this device is currently in the possession of the explanting hospital.This investigation will be updated should further information be provided.
 
Event Description
It was reported that this cardiac resynchronization therapy pacemaker (crt-p) device was found to be in safety mode with limited critical therapy still available.Boston scientific technical services (ts) discussed with boston scientific representatives to determine if the patient is having symptoms of diaphragmatic stimulation or pacing inhibition and recommended device replacement as soon as is reasonable.The device was subsequently explanted and replaced the next day.No patient symptoms or additional adverse patient effects were reported.
 
Event Description
It was reported that this cardiac resynchronization therapy pacemaker (crt-p) device was found to be in safety mode with limited critical therapy still available.Boston scientific technical services (ts) discussed with boston scientific representatives to determine if the patient is having symptoms of diaphragmatic stimulation or pacing inhibition and recommended device replacement as soon as is reasonable.The device was subsequently explanted and replaced the next day.No patient symptoms or additional adverse patient effects were reported.
 
Manufacturer Narrative
Information indicates this device is currently in the possession of the explanting hospital.This investigation will be updated should further information be provided.Upon receipt at our post market quality assurance laboratory, this device was thoroughly inspected and analyzed.Interrogation of the device confirmed it was operating in safety mode due to system resets.It was confirmed that brady therapy remained available.The system resets occurred during a telemetry session and caused the device to enter safety mode.Engineers determined that this device was demonstrating behavior consistent with a high internal cell impedance within the battery.The high internal impedance resulted in the system resets due to temporary high-power consumption related to telemetry attempts and subsequent reversion to safety mode operation.Boston scientific has issued a field safety notice regarding ingenio extended life (el) and cardiac resynchronization therapy pacemaker (crt-p) devices that may exhibit this behavior.This particular device is included in the high impedance in ingenio extended life (el) pacemaker and crt-p advisory population.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INVIVE
Type of Device
IMPLANTABLE DEVICE
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key12620459
MDR Text Key276097389
Report Number2124215-2021-24143
Device Sequence Number1
Product Code NKE
UDI-Device Identifier00802526536632
UDI-Public00802526536632
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030005/S079
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Nurse
Remedial Action Replace
Type of Report Initial,Followup
Report Date 04/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/12/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date06/13/2014
Device Model NumberV173
Device Catalogue NumberV173
Device Lot Number100963
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/01/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/15/2012
Is the Device Single Use? Yes
Type of Device Usage Initial
Removal/Correction NumberZ-2018-2021
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age55 YR
Patient SexFemale
-
-