• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION VALITUDE CRT-P PULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION (CRT-P)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION VALITUDE CRT-P PULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION (CRT-P) Back to Search Results
Model Number U125
Device Problems Signal Artifact/Noise (1036); Failure to Capture (1081); High impedance (1291); Over-Sensing (1438); Pacing Problem (1439); Off-Label Use (1494)
Patient Problem Discomfort (2330)
Event Date 12/01/2018
Event Type  Injury  
Manufacturer Narrative
If additional information is received, a supplemental report will be filed at that time.
 
Event Description
It was reported that this cardiac resynchronization therapy pacemaker exhibited right ventricular (rv) noise with respiration, oversensing, pacing inhibition and loss of capture. It was noted that the left ventricular (lv) lead had previously been placed into the rv port of the device but was since switched back. There was right atrial (ra) loss of capture observed on the non boston scientific ra lead as well. It was noted that the patient had been symptomatic. Subsequently, a revision procedure was performed in which the crt-p was explanted. A new device was successfully implanted and remains in service. No additional adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameVALITUDE CRT-P
Type of DevicePULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION (CRT-P)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key12620473
MDR Text Key276109379
Report Number2124215-2021-28995
Device Sequence Number1
Product Code NKE
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030005/S113
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation
Type of Report Initial,Followup
Report Date 12/19/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/12/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator
Device Expiration Date08/03/2018
Device Model NumberU125
Device Catalogue NumberU125
Device Lot Number703266
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/08/2021
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/25/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/22/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

Patient Treatment Data
Date Received: 10/12/2021 Patient Sequence Number: 1
-
-