BOSTON SCIENTIFIC CORPORATION VALITUDE X4 CRT-P IMPLANTABLE DEVICE Back to Search Results
Model Number U128
Device Problems Signal Artifact/Noise (1036); Over-Sensing (1438); Pacing Problem (1439); Low impedance (2285)
Patient Problem Asystole (4442)
Event Date 04/01/2021
Event Type  malfunction  
Event Description
It was reported that this cardiac resynchronization therapy pacemaker (crt-p) exhibited noise that was being oversensed on the right ventricular (rv) channel, which resulted in the patient experiencing asystole greater than 2 seconds. Additionally, it was noted the pacing impedances dropped a few times however, measurements were within range. Technical services discussed possible cause and provided programming options. At this time, the crt-p and non-boston scientific rv lead remains in service. No additional adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameVALITUDE X4 CRT-P
Type of DeviceIMPLANTABLE DEVICE
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key12620539
MDR Text Key276900357
Report Number2124215-2021-27886
Device Sequence Number1
Product Code NKE
UDI-Device Identifier00802526559402
UDI-Public00802526559402
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030005/S113
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 10/12/2021
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received10/12/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Expiration Date07/10/2017
Device Model NumberU128
Device Catalogue NumberU128
Device Lot Number702656
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received07/30/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/10/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-