The neuro accessory kit was returned to the manufacturer for analysis.The kit was found to be fully functional with no problem found.The reported event could not be duplicated by medtronic personnel.Other relevant device(s) are: product id: 29631, serial/lot #: (b)(4), udi#: (b)(4); product id: 9735571, serial/lot #: unk.If information is provided in the future, a supplemental report will be issued.
|
Medtronic received information regarding a navigation system and automated trajectory guidance unit being used during a catheter placement for a medtronic thermal therapy system procedure.It was reported that there were difficulties placing the bone anchor.It was reported that after alignment with the guidance unit and drilling a burr hole, the bone anchor was not "smooth".It was noted that there were difficulties with alignment of the guidance unit as well.A second burr hole was attempted, but it was noted that the hole was overburred.It was reported that the drape around the guidance unit did not interfere with the guidance unit and that the unit pressed against the mayfield in the room.There was a reported delay to the procedure of less than 1 hour due to this issue.There was no reported impact on patient outcome.
|