• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION INVIVE PULSE GENERATOR, PACEMAKER, WITH CRT-P

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION INVIVE PULSE GENERATOR, PACEMAKER, WITH CRT-P Back to Search Results
Model Number V173
Device Problems Failure to Capture (1081); Pacing Problem (1439); Incorrect, Inadequate or Imprecise Resultor Readings (1535); Use of Device Problem (1670); Impedance Problem (2950)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/11/2021
Event Type  Injury  
Event Description
It was reported that this cardiac resynchronization therapy pacemaker (crt-p) exhibited error codes and was found to be operating in safety mode. A surgical revision was performed to explant and replace the device. During the procedure, all the leads were found to be in the incorrect ports in the device header. The patient's right atrial (ra) lead exhibited no capture, low impedance measurements, increased threshold measurements, and pacing not delivered when required. The device was replaced as planned, and was returned for analysis. No additional adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameINVIVE
Type of DevicePULSE GENERATOR, PACEMAKER, WITH CRT-P
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key12627903
MDR Text Key276382155
Report Number2124215-2021-25897
Device Sequence Number1
Product Code NKE
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030005/S079
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Remedial Action Recall
Type of Report Initial,Followup
Report Date 04/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/13/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator
Device Expiration Date08/04/2014
Device Model NumberV173
Device Catalogue NumberV173
Device Lot Number101582
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/24/2021
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/11/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/08/2012
Is the Device Single Use? Yes
Type of Device Usage Initial

Patient Treatment Data
Date Received: 10/13/2021 Patient Sequence Number: 1
-
-