BAXTER HEALTHCARE CORPORATION PRISMAFLEX M100; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
|
Back to Search Results |
|
Catalog Number 955595 |
Device Problem
Leak/Splash (1354)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 07/28/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
The device was not received for evaluation, however (b)(6) were provided.The visual inspection of the provided (b)(6) observed an external blood leak from the access pod.The cause of the event was supplier issue.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
|
Event Description
|
It was reported that during treatment with a prismaflex m100 set, an internal blood leak was observed; however, inspection of (b)(6) provided of the reported indicated that a minor external blood leak from the access pressure pod had occured.There was no patient injury or medical intervention associated with this event.No additional information is available.
|
|