It was reported that the patient was admitted to the hospital with decompensated heart failure on (b)(6) 2021 and presented to the or (operating room) on (b)(6) 2021 for implant of hm3 (heartmate 3) by a left lateral thoracotomy approach during the index procedure.Prior to going to the or, patient was placed on an intra-aortic balloon pump (iabp) and hm3 was implanted.The transesophageal echocardiogram (tee) noted large amounts of air coming from left ventricle (lv) apex area.Multiple attempts were made to address the bleeding concern and adjust the pump, but there was no resolution to air in the lv.The large amount of air was a concern that the pump was not properly locked and/or possibly damaged.A sterile mini cuff was opened and tested and a small (~1 cm) portion of the locking mechanism was visible on the pump.This exposure was noted before and locking mechanism was not completely flush on left side.(hm3 (b)(4)) was replaced with hm3 (b)(4).Once the second hm3 pump was attached to the initial apical cuff, there was no additional air in the left ventricle.The patient was weaned off bypass, the air appeared to improve as seen on tee.The right ventricle was not doing well and the patient was placed on fem/ fem v-a ecmo (femoral venoarterial extracorporeal membrane oxygenation) to provide additional hemodynamic support using the hm3 as an lv vent.The chest was left open and the patient returned to the icu (intensive care unit).On (b)(6) 2021 , computed tomography (ct) of the head displayed questionable small infarct in the left anterior frontal lobe of uncertain age.Patient began having seizures likely due to air entrainment of first vad (ventricular assist device).The patient passed away on (b)(6) 2021 due to neurological failure to wake up post operation of heartmate 3 implant.Hm3 (b)(4) death report: mfr# 2916596-2021-05725.
|