• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL - NEUROMODULATION (PUERTO RICO, LLC) PROCLAIM¿ 7 ELITE IMPLANTABLE PULSE GENERATOR SCS IPG

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL - NEUROMODULATION (PUERTO RICO, LLC) PROCLAIM¿ 7 ELITE IMPLANTABLE PULSE GENERATOR SCS IPG Back to Search Results
Model Number 3662
Device Problem Wireless Communication Problem (3283)
Patient Problems Failure of Implant (1924); Inadequate Pain Relief (2388)
Event Date 05/07/2021
Event Type  Injury  
Manufacturer Narrative
Date of event is estimated.
 
Event Description
It was reported the patient's ipg failed to establish communication with external devices. A manufacturer representative confirmed the issue and the ipg was deemed inoperable. In turn, surgical intervention may take place at a later date to address the issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NamePROCLAIM¿ 7 ELITE IMPLANTABLE PULSE GENERATOR
Type of DeviceSCS IPG
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION (PUERTO RICO, LLC)
lot a interior - #2 street km 67.5
santana industrial park
arecibo PR 00612
Manufacturer (Section G)
ST. JUDE MEDICAL - NEUROMODULATION (PUERTO RICO, LLC)
lot a interior - #2 street km 67.5
santana industrial park
arecibo PR 00612
Manufacturer Contact
ronnie shalev
6901 preston road
plano, TX 75024
9723098000
MDR Report Key12642929
MDR Text Key276849002
Report Number3006705815-2021-05076
Device Sequence Number1
Product Code LGW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation
Type of Report Initial,Followup
Report Date 11/15/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/15/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator
Device Expiration Date03/07/2021
Device Model Number3662
Device Catalogue Number3662
Device Lot NumberA000076559
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received10/21/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/08/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

Patient Treatment Data
Date Received: 10/15/2021 Patient Sequence Number: 1
-
-