• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION VISIONIST X4 CRT-P PULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION (CRT-P)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION VISIONIST X4 CRT-P PULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION (CRT-P) Back to Search Results
Model Number U228
Device Problem Failure to Capture (1081)
Patient Problem Syncope/Fainting (4411)
Event Date 09/02/2021
Event Type  malfunction  
Event Description
It was reported that after the implant procedure, a manual left ventricular (lv) capture threshold test was performed during normal discharge procedures. However, there was a long pause after the technician finished the test and the patient experienced a syncopal episode. It was alleged the patient lost consciousness due to lack of back-up pacing during manual threshold testing. The patient was discharged with no additional adverse effects and this device remains implanted and in-service.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameVISIONIST X4 CRT-P
Type of DevicePULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION (CRT-P)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key12644187
MDR Text Key276895379
Report Number2124215-2021-29466
Device Sequence Number1
Product Code NKE
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
P030005/S138
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation
Type of Report Initial
Report Date 10/17/2021
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received10/17/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model NumberU228
Device Catalogue NumberU228
Device Lot Number753332
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received09/02/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/03/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device?
Type of Device Usage Initial

-
-