SMITH & NEPHEW, INC. HEALICOIL SA PK 5.5MM W/3 UB-BL CBBL LT; FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE Back to Search Results
Model Number 72203380
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Wound Dehiscence (1154); Unspecified Infection (1930)
Event Date 09/23/2021
Event Type  Injury  
Manufacturer Narrative
Internal complaint reference: case- (b)(4).
 
Event Description
It was reported that, after a lateral instability procedure with a healicoil, one patient experienced wound dehiscence and subsequent infection requiring revision for device replacement.This incident was noticed in a retrospective post-market clinical follow up activity (pmcf) where anonymized data summarizing outcomes were gathered; therefore, additional information is not known and it is not possible to collect it.
 
Manufacturer Narrative
The reported device was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.There was no way to determine if the device contributed to the reported event.The complaint was not confirmed.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.A complaint history review concluded this was an isolated event.Insufficient product identification information was provided and thus a manufacturing record review could not be conducted.Based on the information available, there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.The instructions for use was reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A risk management review found that the reported failure was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate our clinical investigation concluded: the data presented was obtained retrospectively from a post-market clinical follow-up (pmcf) and is anonymized, therefore, further details are not available.Without clinically relevant patient-specific supporting documentation, a thorough medical investigation could not be performed.The root cause and/or patient outcome beyond that which was documented in the article could not be confirmed nor concluded; therefore, no further medical assessment is warranted at this time.Should any additional relevant clinical information be provided, this complaint would be re-assessed.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEALICOIL SA PK 5.5MM W/3 UB-BL CBBL LT
Type of Device
FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer (Section G)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key12648245
MDR Text Key276901865
Report Number1219602-2021-02299
Device Sequence Number1
Product Code MBI
UDI-Device Identifier00885554026313
UDI-Public00885554026313
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
K113294
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Study,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 05/17/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/18/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number72203380
Device Catalogue Number72203380
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received05/12/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age50 YR
Patient SexMale
-
-