• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON DICKINSON INFUSION THERAPY SYSTEMS INC. BD INSYTE-N¿ AUTOGUARD¿ SHIELDED IV CATHETER; INTRAVASCULAR CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON INFUSION THERAPY SYSTEMS INC. BD INSYTE-N¿ AUTOGUARD¿ SHIELDED IV CATHETER; INTRAVASCULAR CATHETER Back to Search Results
Model Number 381511
Device Problems Leak/Splash (1354); Material Split, Cut or Torn (4008)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/15/2021
Event Type  malfunction  
Manufacturer Narrative
Investigation summary: since no samples displaying the condition reported were available for examination, we were unable to fully investigate this incident.Examination of the product involved may provide clarification as to the cause for the reported failure.A device history record review showed no non-conformances associated with this issue during the production of this batch.Information will be captured on trend reports and monitored.Our business team regularly reviews the collected data for identification of emerging trends.
 
Event Description
It was reported when using the bd insyte-n¿ autoguard¿ shielded iv catheter, the device experienced the catheter splitting.This event occurred twice.The following information was provided by the initial reporter.The customer stated: attempted to start iv access.Iv catheter split inversely x2 separate catheters upon insertion into vein.Each occurrence appeared as abnormal rise in skin/subcutaneous tissue causing need to remove catheter to avoid possible infiltration or vein injury with fluid introduction.Re-puncture was necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD INSYTE-N¿ AUTOGUARD¿ SHIELDED IV CATHETER
Type of Device
INTRAVASCULAR CATHETER
Manufacturer (Section D)
BECTON DICKINSON INFUSION THERAPY SYSTEMS INC.
9450 south state street
sandy UT 84070
MDR Report Key12650791
MDR Text Key277250840
Report Number1710034-2021-00891
Device Sequence Number1
Product Code FOZ
UDI-Device Identifier30382903815112
UDI-Public30382903815112
Combination Product (y/n)N
PMA/PMN Number
K952861
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other,user facility
Type of Report Initial
Report Date 09/28/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/18/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date01/31/2024
Device Model Number381511
Device Catalogue Number381511
Device Lot Number1036293
Was Device Available for Evaluation? No
Date Manufacturer Received09/20/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/05/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-