• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COMPANION MEDICAL INC INPEN MMT-105NNPKNA NOVO NORDISK PINK; SYRINGE, PISTON

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COMPANION MEDICAL INC INPEN MMT-105NNPKNA NOVO NORDISK PINK; SYRINGE, PISTON Back to Search Results
Model Number MMT-105NNPKNA
Device Problems Break (1069); Computer Software Problem (1112)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/01/2021
Event Type  malfunction  
Manufacturer Narrative
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event Description
The customer reported via phone call that there was an inaccuracy between the dose intended and dose recorded.The dose was greater than 1.0 unit of insulin.Customer also reported insulin pen was damaged.No harm requiring medical intervention was reported.The insulin pen will not be returned for analysis.
 
Manufacturer Narrative
Customer reports: inpen app is not recording the exact doses dialed on the inpen.The app is always recording a half unit or 1 unit less than what i dispensed.Serial number: (b)(6).Software version: 3.8.5.Color: pink.Battery life remaining: <11 months.Per visual inspection: the alignment of the dose indicator mark and printed values on the dose knob is nominal.No misalignment of the dial was noted.No physical damage to injection foot or inpen was noted.The customer complaint of misalign dial was not confirmed.Unit paired successfully to commercial app.My inpen menu displayed: the following test values were dialed and dosed: 4.0u, 4.0u, 4.0u, 4.0u,16.0u, 5.0u, 5.0u, 30.0u, 30.0u and 30.0u.All values displayed accurately in the logbook.All doses were written to the app logbook within 1 minute.The screw was not bent, advanced when dosage knob was pressed dialing a dosage and retracted appropriately.No resistance was observed when dosing without a cartridge installed.The screw advanced every time 30.0u was dialed and dosed until the screw reached max extension.In conclusion: the patient complaint of dose log inaccuracies could not be confirmed.Inpen did not produce any false dose logs during investigation.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INPEN MMT-105NNPKNA NOVO NORDISK PINK
Type of Device
SYRINGE, PISTON
Manufacturer (Section D)
COMPANION MEDICAL INC
1223 world trade drive, 100
san diego CA 92128
Manufacturer (Section G)
COMPANION MEDICAL INC
1223 world trade drive, 100
san diego CA 92128
Manufacturer Contact
tricha miles
1223 world trade drive, 100
san diego, CA 92128
7635140379
MDR Report Key12656694
MDR Text Key279781452
Report Number3012822846-2021-00789
Device Sequence Number1
Product Code FMF
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 11/09/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/19/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date03/01/2023
Device Model NumberMMT-105NNPKNA
Device Catalogue NumberMMT-105NNPKNA
Device Lot NumberB94WT
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/20/2021
Is the Reporter a Health Professional? No
Date Manufacturer Received11/02/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age73 YR
-
-