• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. URETERO-RENO FIBERSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. URETERO-RENO FIBERSCOPE Back to Search Results
Model Number URF-P7
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The scope was returned to the olympus' regional repair center (rrc) for evaluation.An air leak was noted at the scope¿s biopsy channel.The distal end cap was found burnt.The insertion tube was noted to be broken.Paint was noted to be peeling off the control grip unit.Light guide bundle breakage was noted causing uneven illumination.An investigation is ongoing to obtain additional information regarding the reported event.If additional information is received this report will be supplemented accordingly.
 
Event Description
The customer reported that after the conclusion of an unspecified procedure, the sheath detached from the scope¿s tip.There was no patient injury reported.
 
Manufacturer Narrative
This report is being supplemented to provide a correction for g2 and h4.This report also includes additional information based on the legal manufacturer's final investigation.G2 - checked 'other' to add the country italy.H4 - corrected device manufacturer date to november 8, 2018.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the cause of the sheath detaching from the scope tip likely occurred from user handling the device during the procedure.The specific root cause of the user handling could not be determined at this time.The following information is stated in the instructions for use regarding proper handling of the device which may have prevented the event: "preparation and inspection - inspection of the endoscope: inspect the entire distal end of the endoscope including the objective lens and examination light outlet for any irregularities such as scratches, chips, cracks, stains, discoloration, deformation, and gaps around the lens.Operation - using the laser system: warning: do not irradiate the laser without viewing the tip of the laser probe and the aiming beam in the endoscopic image.This may cause patient injury, and the instrument channel may be damaged.Do not withdraw the laser probe while irradiating the laser.Patient injury and/or endoscope damage may occur." olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
URETERO-RENO FIBERSCOPE
Type of Device
URETERO-RENO FIBERSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key12658125
MDR Text Key277211341
Report Number8010047-2021-13350
Device Sequence Number1
Product Code FGB
UDI-Device Identifier04953170403811
UDI-Public04953170403811
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
K181451
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,User Facility
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 12/20/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/19/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberURF-P7
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/27/2021
Was the Report Sent to FDA? No
Date Manufacturer Received12/16/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/08/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-