Model Number TPL0059 |
Device Problems
Imprecision (1307); Mechanical Problem (1384)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/27/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Other relevant device(s) are: product id: asm0206.A medtronic representative went to the site to test the equipment.Testing revealed that the surgical arm was replaced.The system then passed the system checkout and was found to be fully functional.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Medtronic received information regarding a guidance system being used during a spinal procedure.It was reported that there were deviations during a l4-l5 stage 2 ap surgery post alif.The surgical system was mounted to the patient using a schanz pin placed in the left psis connected to a bone mount bridge.The surgeon started at l4 and then moved to l5 when placing screws.Navigation and drill were used during the procedure.The manufacturer representative was not sure if navigation accuracy was checked during the procedure.An accuracy check was not completed prior to the procedure.The l5 screws were deviated to a more cephalad position by 3.5 to 5 mm.The screw violated an adjacent endplate.The lateral to medial screws were on trajectory.The c-arm was used to confirm the placement of the screws during the procedure and that was when the deviation was found.An o-arm spin was done to confirm placement.There were no issues with the guidance system during the procedure.The cause of the deviation was not determined.The screws were repositioned during the procedure using flouro.There was no patient harm intra-op and post-op.Neuromonitoring was used and the screws tested safe.There was no patient harm and the procedure was delayed less than an hour.
|
|
Manufacturer Narrative
|
Section d information references the main component of the system.Other relevant device(s) are: product id: asm0206, serial/lot #: (b)(6).H3: analysis of the software exports and logs found the complaint was confirmed.Clinical export data file was thoroughly inspected.The log file was examined with respect to all intraoperative fluoro images in order to inspect and understand procedure workflow.Fluoro images were checked and 3d registration was attempted with the registration 3d marker images utilized during the operation on a mazor x r<(>&<)>d workstation.Analysis reviewed the planning for the case and did not find any issues related to the planning.The nature of the deviation was reviewed.Post op images provided show that both l4 and l5 screws executed deviated superiorly.Analysis concluded the probable cause of the deviation experienced in the or was compromised accuracy caused by a mechanical damage in the surgical arm.The fact that replacing the arm solved the issues suggest this was the cause of the mentioned deviation.Analysis results of the surgical arm were not available as of the date of this report.A follow up report will be submitted when analysis is complete.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
The surgical arm was returned for analysis.Analysis confirmed failures with the surgical arm.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Additional information received from a manufacturer representative reported that a surgical arm accuracy check was done as part of the pre-op arm set up.An advanced accuracy test and intra-op accuracy verification using the divot and reference frame was not done.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Additional information received from a manufacturer representative reported that the case was minimally invasive and a drill was used during the procedure.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
H3: analysis of the surgical arm found the complaint was confirmed.Visual/physical examination and functional testing found the surgical arm failed an accuracy test, the j6 communication harness had a problem, the front plate was missing, and the j2/j4 plastic covers were broken.The surgical arm was repaired, recalibrated and passed all testing.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|