Model Number U225
Device Problem High impedance (1291)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/08/2021
Event Type  malfunction  
Event Description
It was reported that this cardiac resynchronization therapy pacemaker (crt-p), and non-boston scientific right atrial (ra) lead exhibited intermittent pace impedance measurements high out of range, greater than 3,000 ohms. Technical services (ts) discussed troubleshooting and programming options. The device was reprogrammed for optimization. No adverse patient effects were reported. The device remains in service.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameVISIONIST CRT-P
Type of DevicePULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION (CRT-P)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key12660796
MDR Text Key277266822
Report Number2124215-2021-30172
Device Sequence Number1
Product Code NKE
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030005/S138
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation
Type of Report Initial
Report Date 10/19/2021
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received10/19/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Expiration Date04/18/2021
Device Model NumberU225
Device Catalogue NumberU225
Device Lot Number712837
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received09/14/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/30/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device?
Type of Device Usage Initial

-
-