Reportable based on device analysis completed on 14oct2021.It was reported that the device was kinked.The 32mmx3.0mm, 85% stenosed target lesion was located in the moderately tortuous and moderately calcified left anterior descending artery.A guidezilla ii catheter-guide extension was selected for use.During the procedure, it was noted that the physician found the interface of transition segment was kinked.The device was directly removed and the procedure was completed with another of the same device.There were no complications reported and the patient is stable.However, device investigations revealed that a collar damage which created a hole or tear in the shaft.
|