Model Number 3662 |
Device Problem
Wireless Communication Problem (3283)
|
Patient Problems
Failure of Implant (1924); Inadequate Pain Relief (2388)
|
Event Date 08/02/2021 |
Event Type
Injury
|
Manufacturer Narrative
|
Date of event is estimated.
|
|
Event Description
|
It was reported the patient had an unrelated shoulder surgery in (b)(6) 2021 where the device was not placed in surgery mode.Subsequently, the patient has been unable to connect to the ipg and the patient controller is displaying an error message since the procedure.Surgical intervention may take place at a later date.
|
|
Manufacturer Narrative
|
An inoperable implant was reported to abbott.The system was not set to surgery mode while the patient underwent an unrelated surgery where electro-cautery may have been used.The implant was not returned for analysis.A review of documentation supplied with the implant states that electro-surgery devices should not be used in close proximity to an implanted system.Based on the information received, the cause of the reported incident is consistent with user error.
|
|
Event Description
|
Additional information received indicates the ipg was explanted and replaced to address the issue.
|
|
Manufacturer Narrative
|
During processing of this complaint, attempts were made to obtain complete patient information.
|
|
Manufacturer Narrative
|
An inoperable implant was reported to abbott.The system was not set to surgery mode while the patient underwent an unrelated surgery where electro-cautery may have been used.The implant was not returned for analysis.A review of documentation supplied with the implant states that electro-surgery devices should not be used in close proximity to an implanted system.Based on the information received, the cause of the reported incident is consistent with user error.
|
|