Model Number N/A |
Device Problem
Fracture (1260)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/21/2021 |
Event Type
Injury
|
Manufacturer Narrative
|
(b)(4).Report source: (b)(6).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.Has not been returned yet.
|
|
Event Description
|
It was reported that the bone depth gauge front piece broke into patients finger.Attempts have been made and there is no further information at this time.
|
|
Event Description
|
No further event information available at the time of this report.
|
|
Manufacturer Narrative
|
This follow-up report is being submitted to relay additional information.The following sections were updated: b4; b5; g3; g6; h1; h2; h3; h4; h6; h10 visual examination of the provided picture identified the device shows slight wear; scratches and nicks present throughout.Fracture cannot be confirmed as the fragment is not present in the image and the tip of the device is not shown.Part number verified, but lot number is not visible in the provided picture.The dhr was reviewed and no discrepancies relevant to the reported event were found.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|