• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: KAVO DENTAL GMBH INTRA SURGICAL HANDPIECE 3610N2; DENTAL SURGICAL HANDPIECE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

KAVO DENTAL GMBH INTRA SURGICAL HANDPIECE 3610N2; DENTAL SURGICAL HANDPIECE Back to Search Results
Model Number 3610 N2
Device Problems Mechanical Problem (1384); Overheating of Device (1437)
Patient Problem Burn(s) (1757)
Event Date 09/03/2021
Event Type  malfunction  
Manufacturer Narrative
The test run prior to the repair showed that the small ball bearing in the front of the handpiece was worn out.This bearing guides the clamped tool and is therefore loaded during use.A worn bearings leads to high friction and hence to heat up.To avoid such issues the instruction for use contains already several warnings and notes how to prepare and use the product: technical condition: a damaged device or components could injure patients, users and third parties.Only operate devices or components if they are undamaged on the outside.Check that the device is working properly and is in satisfactory condition before each use.If there are any broken or damaged parts or clearly visible changes on the surface, the parts must be checked by the service.Safety checks may only be performed by trained service personnel.If the following defects occur, stop working and have the service personnel carry out repair work: malfunctions, damage, irregular running noise, excessive vibration, overheating, dental bur or diamond grinder are not firmly locked in the handpiece.Kavo recommends specifying in-house service intervals where the medical device is brought to a professional shop for cleaning, servicing and a function check.Define the service interval depending on the frequency of use.
 
Event Description
During a dental surgical intervention, the handpiece heated up and caused a burn inside the patient's lip.Patient was fine after the treatment.There is a follow-up scheduled in two weeks related to the surgical intervention.Related to the burn there is no medical care necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTRA SURGICAL HANDPIECE 3610N2
Type of Device
DENTAL SURGICAL HANDPIECE
Manufacturer (Section D)
KAVO DENTAL GMBH
bismarckring 39
biberach / riss, bw 88400
GM  88400
Manufacturer (Section G)
KAVO DENTAL GMBH
bismarckring 39
biberach / riss, bw 88400
GM   88400
Manufacturer Contact
klaus reisenauer
bismarckring 39
biberach / riss, 88400
GM   88400
MDR Report Key12671888
MDR Text Key277630567
Report Number3003637274-2021-00038
Device Sequence Number1
Product Code EFB
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K934783
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Dentist
Type of Report Initial
Report Date 10/21/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/21/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number3610 N2
Device Catalogue Number0.524.5610
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/23/2021
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/23/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/23/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-