Model Number 466P306X
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Perforation (2001)
Event Date 08/30/2021
Event Type  Injury  
Manufacturer Narrative
As reported, the patient underwent placement of a trapease vena cava filter.The indication for the filter placement was not reported.At some point after the filter implantation, the patient became aware that the filter was associated with perforation.The product was not returned for analysis and the sterile lot number has not been provided; therefore, no device analysis nor device history record review could be performed.The trapease vena cava filter is indicated for use in the prevention of recurrent pulmonary embolism (pe) via percutaneous placement in the inferior vena cava (ivc) for patients in which anticoagulants are contraindicated, anticoagulant therapy for thromboembolic disease has failed, emergency treatment following massive pe where anticipated benefits of conventional therapy are reduced or for chronic, recurrent pe where anticoagulant therapy has failed, or is contraindicated.The purpose of a vena cava filter is to catch thrombus from the lower extremities as it travels along normal blood flow patterns up towards the heart.Without images or procedural films for review, the reported perforation could not be confirmed and the exact cause could not be determined.It is unknown if the tilt contributed to the reported perforation.A review of the instructions for use (ifu) notes vessel damage such as intimal tears and perforation as procedural complications related to ivc filters.Studies have noted a greater than 80% perforation rate overall, with all filters imaged after 71 days from implantation revealing some level of perforation.Clinical factors that may have influenced the event include the patient¿s pre-existing co-morbidities, pharmacological issues and lesion characteristics.Given the limited information available for review, there is nothing to suggest that a malfunction in the design and manufacturing process of the device; therefore, no corrective action will be taken.Should additional information become available, the file will be updated accordingly.
 
Event Description
As reported by the legal department, the patient underwent placement of a trapease vena cava filter.The filter subsequently malfunctioned and caused injury and damages to the patient including, but not limited to, perforation.As a direct and proximate result, the patient suffered life-threatening injuries and damages and required extensive medical care and treatment.As a further proximate result, the patient has suffered and will continue to suffer significant medical expenses, pain and suffering, and other damages.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN TRAPEASE
Type of Device
FILTER, INTRAVASCULAR, CARDIOVASCULAR
Manufacturer (Section D)
CORDIS CORPORATION
14201 nw 60th avenue
miami lakes FL 33014
Manufacturer (Section G)
CORDIS CORPORATION
14201 nw 60th avenue
miami lakes FL 33014
Manufacturer Contact
karla castro
14201 nw 60th avenue
miami lakes, FL 33014
7863138372
MDR Report Key12672292
MDR Text Key277632797
Report Number9616099-2021-04967
Device Sequence Number1
Product Code DTK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K020316
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other
Reporter Occupation Other
Type of Report Initial
Report Date 10/21/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/21/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number466P306X
Device Catalogue Number466P306X
Was Device Available for Evaluation? No
Date Manufacturer Received09/30/2021
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Life Threatening;
-
-