DATASCOPE CORP. - MAHWAH CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 220V; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number 0998-00-3023-55
Device Problems Inability to Auto-Fill (1044); Device Contamination with Body Fluid (2317)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/27/2021
Event Type  malfunction  
Manufacturer Narrative
Testing of actual/suspected device: a getinge field service engineer (fse) was dispatched to investigate.The fse evaluated the iabp unit and observed blood stains in the crm assembly, as well as the internal tubing, purge assembly, safety disk, blood detect glass, and other related tubing connections.The customer has been presented with a cost estimate to repair the iabp unit.Repairs are pending approval.A supplemental report will be submitted when additional information is provided.
 
Event Description
It was reported that prior to use, the cs300 intra-aortic balloon pump (iabp) alarmed autofill failure, as well as blood stains found in the internal tubing and on the drain port of the crm assembly.There was no patient involvement, and no adverse event reported.
 
Manufacturer Narrative
Updated fields: b4, g3, g6, g7, h2, h4, h6 (type of investigation, investigation findings, investigation conclusions), h10.
 
Event Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 220V
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key12672541
MDR Text Key278217473
Report Number2249723-2021-02403
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567112541
UDI-Public10607567112541
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
K063525
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 07/08/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/21/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0998-00-3023-55
Device Catalogue Number0998-00-3023-55
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/08/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/22/2019
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-