Model Number 97713
Device Problems High impedance (1291); Patient Device Interaction Problem (4001)
Patient Problem Inadequate Pain Relief (2388)
Event Date 10/13/2021
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received from a manufacturer representative regarding a patient who was implanted with an implantable neurostimulator (ins) for non-malignant pain. It was reported that the impedances results showed many values are >10000ohms and some are in the 600ohm range. Pair 8 1 was 620ohms. Patient is using therapy for neck and shoulder pain. The caller indicated that therapy works "mostly well" but the patient feels an increased stimulation sensation when they move their neck. The caller indicated that the shoulder pain has also been getting progressively worse related to playing hockey and the md ordered xrays to further evaluate the shoulder. The caller was not with the patient. Tss reviewed impedance information with the caller. The caller indicated that they would follow-up with the md.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameSURESCAN
Type of DeviceSTIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
glen belmer
7000 central avenue ne rcw215
minneapolis, MN 55432
6122713209
MDR Report Key12673752
MDR Text Key277721517
Report Number3004209178-2021-15678
Device Sequence Number1
Product Code LGW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P840001
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation
Type of Report Initial
Report Date 10/21/2021
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received10/21/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator
Device Expiration Date02/28/2018
Device Model Number97713
Device Catalogue Number97713
Was Device Available for Evaluation? No
Is the Reporter a Health Professional?
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received10/13/2021
Was Device Evaluated by Manufacturer? No Answer Provided
Date Device Manufactured01/30/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-