DATASCOPE CORP. - MAHWAH CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 110V; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
|
Back to Search Results |
|
Model Number 0998-00-3023-53 |
Device Problem
Erratic or Intermittent Display (1182)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/29/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Device not accessible for testing: at this time, the customer has not requested getinge to evaluate the iabp unit involved in this event.Attempts to communicate with the customer have been made to obtain further repair information.A supplement report will be submitted should additional information be provided.Device not returned to manufacturer.
|
|
Event Description
|
It was reported that during a routine check, a rental cs300 intra-aortic balloon pump (iabp) from agiliti had intermittent display issues upon power-up.It was also reported that the cs300 will be evaluated once returned to agiitii warehouse.There was no patient involvement, and no adverse event reported.
|
|
Manufacturer Narrative
|
Testing of actual/suspected device (10): a getinge field service engineer (fse) was dispatched to investigate.The fse evaluated the iabp unit, and was unable to reproduce the reported issue.The fse performed a full preventive maintenance (pm) with all calibration, functional and safety checks to meet factory specifications.Unit passed all calibration, functional and safety test per factory specifications.The iabp was then released to the customer and cleared for clinical service.A supplemental report will be submitted upon completion of our investigation.
|
|
Event Description
|
N/a.
|
|
Event Description
|
N/a.
|
|
Manufacturer Narrative
|
Updated fields: b4, e1 (initial reporter name), e2, e3, g3, g6, h2, h4, h6, h10, h11.Corrected fields: e1 (telephone number), g1.Initial reporter's additional information: avp of cv services.
|
|