MAUDE Adverse Event Report: ARROW INTERNATIONAL INC. ULTRAFLEX IAB: 7.5FR 30CC SYSTEM, BALLOON, INTRA-AORTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARROW INTERNATIONAL INC. ULTRAFLEX IAB: 7.5FR 30CC SYSTEM, BALLOON, INTRA-AORTIC Back to Search Results
Model Number IPN000257
Device Problem Fluid Leak (1250)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/23/2021
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that the staff was getting high-pressure alarms. The alarms were happening consistently, and the user was unable to make the pump work. It was noted that the patient was not moving around, did not have a steep angle of insertion, and there were no visible kinks. When the staff checked for blood in the driveline, the staff confirmed there were dark flecks on the inside of the tubing. As a result, the iab was removed, and patient is stable with no signs of any embolic event. There was no report of patient complications, serious injury or death.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameULTRAFLEX IAB: 7.5FR 30CC
Type of DeviceSYSTEM, BALLOON, INTRA-AORTIC
Manufacturer (Section D)
ARROW INTERNATIONAL INC.
reading PA
Manufacturer (Section G)
ARROW INTERNATIONAL INC.
16 elizabeth drive
chelmsford MA 01824
Manufacturer Contact
kaylia pen
16 elizabeth drive
chelmsford, MA 01824
MDR Report Key12683217
MDR Text Key277953802
Report Number3010532612-2021-00314
Device Sequence Number1
Product Code DSP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K000729
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation
Type of Report Initial,Followup
Report Date 09/23/2021
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received10/22/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model NumberIPN000257
Device Catalogue NumberIAB-06830-U
Device Lot Number18F21D0042
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/11/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/29/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-