• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES GMBH ELECTRIC PEN DRIVE 60,000 RPM; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH ELECTRIC PEN DRIVE 60,000 RPM; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT Back to Search Results
Catalog Number 05.001.010
Device Problem Unintended System Motion (1430)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/01/2021
Event Type  malfunction  
Manufacturer Narrative
Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Additional narrative: the actual device was returned for evaluation.During repair, an evaluation was performed and it was determined that the reported condition was confirmed.The assignable root cause was determined to be due to component wear.Udi: (b)(4).
 
Event Description
It was reported that the pen drive device was returned for service.During an in-house engineering evaluation, it was determined that the device ran in unexpected direction/mode, had unintended activation/motion and failed pre-test for check for unintended motion.It was not reported if the device was used in surgery, or if there was patient involvement.It was not reported if there were any delays in the a surgical procedure, or if a spare device was available for use.It was not reported if there were any injuries, medical intervention or prolonged hospitalization.The exact date of this event was unknown but was noted to have occurred in 2021.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELECTRIC PEN DRIVE 60,000 RPM
Type of Device
INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
SYNTHES PRODUKTIONS GMBH
hauptstrasse 24
waldenburg 4437
SZ   4437
Manufacturer Contact
kara ditty-bovard
eimattstrasse 3
oberdorf 4436
SZ   4436
6107428552
MDR Report Key12688854
MDR Text Key278118483
Report Number8030965-2021-08927
Device Sequence Number1
Product Code HWE
UDI-Device Identifier07611819167711
UDI-Public07611819167711
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other
Type of Report Initial
Report Date 10/16/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/25/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number05.001.010
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/08/2021
Date Manufacturer Received10/15/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/24/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-