Model Number 0684-00-0469-01 |
Device Problem
Calibration Problem (2890)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/05/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Complete event site name: (b)(6).The device will not be returned to the manufacturer so we are unable to complete an evaluation.If provided we will send a supplemental report with our additional findings.Complaint record id # (b)(4).
|
|
Event Description
|
It was reported that after less than 12 hours of intra-aortic balloon (iab) therapy, the console generated an unable to calibrate alarm.The insertion was reported to be axillary, which is not the method described in the device instructions for use.There was no patient harm or adverse event reported.
|
|
Manufacturer Narrative
|
The device has not been returned to the manufacturer so we are unable to complete an evaluation.Analysis of production (3331/213) - the device history records review concluded that there were no ncmrs, rework, or deviations documented for the reported lot/serial number.Based on the review results, it was determined that there is no relation between the manufacturing process and the reported failure.Historical data analysis (4109/213) - the review of the historical data was performed.Trend analysis (4110/213) - the overall complaint trend data for the period oct-2019 to sep-2021 was reviewed.Communication/interviews: (4111/213) communication/interviews were performed to obtain all possible information.Reference complaint #(b)(4).
|
|