• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES GMBH COMPACT AIR DRIVE II; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH COMPACT AIR DRIVE II; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT Back to Search Results
Catalog Number 511.701
Device Problem Physical Resistance/Sticking (4012)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/01/2021
Event Type  malfunction  
Event Description
It was reported by (b)(6) that during service and evaluation, it was determined that the trigger of the compact air drive device was sticky, and the device had an insufficient/ low power.It was further determined that the device failed pretest for check for sticky trigger, check function of device and check power with power test bench.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of the event was unknown.However, it was reported that the event occurred in 2021.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Manufacturer Narrative
Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.Device evaluation: the actual device was returned for evaluation.During repair, an evaluation was performed and it was determined that the reported condition of a sticky trigger identified during service and evaluation was confirmed.The assignable root cause was determined to be due to component failure from wear.Udi: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COMPACT AIR DRIVE II
Type of Device
INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
MDR Report Key12699361
MDR Text Key285400527
Report Number8030965-2021-08993
Device Sequence Number1
Product Code HWE
UDI-Device Identifier07611819085268
UDI-Public07611819085268
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial
Report Date 10/21/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/26/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number511.701
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/21/2021
Date Manufacturer Received10/21/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-