DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID, TYPE B PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
|
Back to Search Results |
|
Model Number 0998-00-0800-53 |
Device Problems
Pumping Stopped (1503); Unexpected Shutdown (4019)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/05/2021 |
Event Type
malfunction
|
Event Description
|
It was reported that during use on a patient the cardiosave intra-aortic balloon pump (iabp) was shutting off randomly without any warning.No patient harm, serious injury or adverse event was reported.
|
|
Manufacturer Narrative
|
Additional information is being requested in regards to whether or not the customer has requested for getinge to service the iabp unit involved.A supplemental report will be submitted when additional information is provided.
|
|
Manufacturer Narrative
|
Testing of actual/suspected device (10/3233): a getinge field service engineer (fse) evaluated the iabp unit, but was unable to reproduce the reported issue.The logs showed no signs of power failure, the system checkout was completed, however the systems safety disk expired due to inflations and the pump was taken out of service until parts are available for install.Another getinge fse arrived at the customer's site at a later date to perform preventative maintenance (pm).The fse replaced the expired 9v battery, both battery packs, and the tidal volume disk.All parts replaced were in relation to pm service and not the reported issue.Subsequently, the fse completed pm service including all safety, functionality, and calibration checks and all tests passed to factory specifications.The iabp unit was cleared for clinical use and released to the customer.A supplemental report will be submitted upon completion of our investigation.
|
|
Manufacturer Narrative
|
Updated fields: b4, g4, g7, g8, h2, h4, h6(investigation type, investigation findings & investigation conclusions), h10, h11 corrected fields: g1(contact person).
|
|