• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DATASCOPE CORP. - MAHWAH CS100 INTRA-AORTIC BALLOON PUMP, ENGLISH, 220V; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE CORP. - MAHWAH CS100 INTRA-AORTIC BALLOON PUMP, ENGLISH, 220V; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number 0998-00-3013-55
Device Problems Failure to Zero (1683); Appropriate Term/Code Not Available (3191)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/06/2021
Event Type  malfunction  
Manufacturer Narrative
A getinge field service engineer (fse) was dispatched to investigate.The fse evaluated the cs100 intra-aortic balloon pump, refurbished, english,non-uts,dom iabp unit and discovered after diagnosis that the problem was in the pressure transducer cable, which had gone defective and needs to be replaced.A supplemental report will be submitted when additional information is provided.
 
Event Description
It was reported that the cs100 unit was unable to zero pressure.It is unknown the circumstances under which the event occurred.It is also unknown if there was a patient involvement.However, there was no adverse event reported.
 
Event Description
It was reported that the cs100 unit was unable to zero pressure.There was no patient involvement reported.
 
Manufacturer Narrative
A getinge field service engineer (fse) evaluated the unit for the reported malfunction.The fse determined that the pressure transducer cable was defective and required replacement.The customer repaired the unit by replacing the pressure transducer cable.The unit was then working properly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CS100 INTRA-AORTIC BALLOON PUMP, ENGLISH, 220V
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key12707739
MDR Text Key281420668
Report Number2249723-2021-02479
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567107295
UDI-Public10607567107295
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
K031636
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup,Followup
Report Date 08/11/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/27/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0998-00-3013-55
Device Catalogue Number0998-00-3013-55
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/11/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/02/2010
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
UNKNOWN.
-
-