Model Number 0998-UC-3013-53
Device Problem Appropriate Term/Code Not Available (3191)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/06/2021
Event Type  malfunction  
Manufacturer Narrative
A getinge field service engineer (fse) was dispatched to investigate. The fse evaluated the cs100 intra-aortic balloon pump, refurbished, english,non-uts,dom iabp unit and discovered after diagnosis that the problem was in the pressure transducer cable, which had gone defective and needs to be replaced. A supplemental report will be submitted when additional information is provided.
 
Event Description
It was reported that the cs100 unit was unable to zero pressure. It is unknown the circumstances under which the event occurred. It is also unknown if there was a patient involvement. However, there was no adverse event reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameCS100 INTRA-AORTIC BALLOON PUMP, REFURBISHED, ENGLISH,NON-UTS,DOM
Type of DeviceSYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
MDR Report Key12707739
MDR Text Key281420668
Report Number2249723-2021-02479
Device Sequence Number1
Product Code DSP
Combination Product (y/n)N
PMA/PMN Number
K031636
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial
Report Date 10/25/2021
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received10/27/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number0998-UC-3013-53
Device Catalogue Number0998-UC-3013-53
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received10/06/2021
Is This a Reprocessed and Reused Single-Use Device?

-
-