Model Number VNL-1570STK |
Device Problem
Corroded (1131)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/08/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
If additional information becomes available, a supplemental report will be filed with the new information.
|
|
Event Description
|
The lg plug was found to be corroded and colour fading on (b)(6) 2020.Then changing the light guide plug.This hospital reported the same defect on (b)(6) 2020.No scratch on the scope connection of the processor but the aging level is serious.The distributor applied for spare part: scope connection assy and lg plug for this hospital.This event occurred at the time of after use.There was no report of patient harm.
|
|
Manufacturer Narrative
|
Correction information g6: follow up #1 h2:if follow-up, what type? h3:device evaluated by manufacture h6: coding changed based on the investigation result additional information h4:device manufacture date evaluation summary the user used the peracetic acid which is not recommended chemistry solution which the reason cause the problem.Pentax replaced the new lg plug.
|
|