• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON DICKINSON INFUSION THERAPY SYSTEMS INC. BD INSYTE¿ AUTOGUARD¿ BC SHIELDED IV CATHETER; INTRAVASCULAR CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON INFUSION THERAPY SYSTEMS INC. BD INSYTE¿ AUTOGUARD¿ BC SHIELDED IV CATHETER; INTRAVASCULAR CATHETER Back to Search Results
Catalog Number 381023
Device Problem Retraction Problem (1536)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/28/2021
Event Type  malfunction  
Manufacturer Narrative
A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event Description
It was reported when using the bd insyte¿ autoguard¿ bc shielded iv catheter the needle would not retract.The following information was provided by the initial reporter.The customer reported as follows: "after venipuncture, the hcp pressed the button to activate the safety mechanism; however, the needle was not retracted.".
 
Event Description
It was reported when using the bd insyte¿ autoguard¿ bc shielded iv catheter the needle would not retract.The following information was provided by the initial reporter.The customer reported as follows: "after venipuncture, the hcp pressed the button to activate the safety mechanism; however, the needle was not retracted.".
 
Manufacturer Narrative
H6: investigation summary our quality engineer inspected the sample and photographs submitted for evaluation.Bd received one unit and four photos.Upon inspection of the received unit and the photos, it was identified that the button had been pressed but the needle had not retracted.The reported defect was confirmed.Further inspection revealed that adhesive was present between the hub and grip, preventing the hub from rotating.The adhesive had adhered the hub and grip together preventing retraction from occurring.During manufacturing, adhesive may be incorrectly placed due to adhesive build up or part misalignment.Preventative maintenance and in process sampling are performed at regular intervals to mitigate the risk from this type of defect.A device history record review showed no non-conformances associated with this issue during the production of this batch.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD INSYTE¿ AUTOGUARD¿ BC SHIELDED IV CATHETER
Type of Device
INTRAVASCULAR CATHETER
Manufacturer (Section D)
BECTON DICKINSON INFUSION THERAPY SYSTEMS INC.
9450 south state street
sandy UT 84070
Manufacturer (Section G)
BECTON DICKINSON INFUSION THERAPY SYSTEMS INC.
9450 south state street
sandy UT 84070
Manufacturer Contact
katie swenson
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key12709281
MDR Text Key278824446
Report Number1710034-2021-00922
Device Sequence Number1
Product Code FOZ
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K201075
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 11/05/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/28/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date11/30/2023
Device Catalogue Number381023
Device Lot Number0337822
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/04/2021
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/05/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/03/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-