Catalog Number 383033
Device Problem Leak/Splash (1354)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/13/2021
Event Type  malfunction  
Manufacturer Narrative
Investigation summary: in response to the event reported by your facility a device history review was conducted for lot number 1051521. According to the sampling plan applied for product performance, this lot was accepted and released without defects being noted during the final assembly or visual inspections. A sample could not be obtained for evaluation and testing; in lieu of the affected device, functional testing was performed on retention samples for this lot, the results of these show that the tested units performed within product specifications. Unfortunately without the ability to investigate the affected unit our quality engineers were unable to determine the root cause for this complaint.
 
Event Description
It was reported when using the bd intima-ii¿ closed iv catheter system there was leakage. The following information was provided by the initial reporter. The customer stated: "the infusion device was connected for infusion. About 10 minutes later, leakage of liquid medicine was found at the clamp. ".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameBD INTIMA-II¿ CLOSED IV CATHETER SYSTEM
Type of DeviceINTRAVASCULAR CATHETER
Manufacturer (Section D)
BECTON DICKINSON MEDICAL DEVICES CO., LTD. SUZHOU
no.1 liangpu street
suzhou industrial park
suzhou 21512 6
CH 215126
MDR Report Key12709918
MDR Text Key284291147
Report Number3014704491-2021-00229
Device Sequence Number1
Product Code FOZ
Combination Product (y/n)N
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,other,user facility
Type of Report Initial
Report Date 10/07/2021
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received10/28/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number383033
Device Lot Number1051521
Was Device Available for Evaluation? No
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received09/29/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/20/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-