• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHILIPS NORTH AMERICA LLC HEART START XL; DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHILIPS NORTH AMERICA LLC HEART START XL; DEFIBRILLATOR Back to Search Results
Model Number M4735A
Device Problem Naturally Worn (2988)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/06/2021
Event Type  malfunction  
Event Description
It was reported to philips that the device's control panel was blurred.There was no patient involvement.
 
Manufacturer Narrative
Complaint evaluation: the customer requested that a philips field service engineer (fse) be dispatched to the customer site.Upon evaluation of the device, the reported issue was confirmed and the cause was traced to a faulty bezel assembly.Customer resolution and conclusion: based on the conclusion of the evaluation, it was determined that this was a malfunction of the bezel assembly.The part was replaced and the device passed all performance assurance testing.The device remains at the customer site.There is no indication of a systemic problem.No further investigation or action is warranted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEART START XL
Type of Device
DEFIBRILLATOR
Manufacturer (Section D)
PHILIPS NORTH AMERICA LLC
22100 bothell everett highway
bothell WA 98021
Manufacturer (Section G)
PHILIPS NORTH AMERICA LLC
22100 bothell everett highway
bothell WA 98021
Manufacturer Contact
dana tackett
22100 bothell everett highway
bothell, WA 98021
9095703538
MDR Report Key12711307
MDR Text Key278781885
Report Number3030677-2021-15260
Device Sequence Number1
Product Code MKJ
Combination Product (y/n)N
Reporter Country CodeTW
PMA/PMN Number
K021453
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/28/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberM4735A
Device Catalogue NumberM4735A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/04/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/15/2006
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-